OBJECTIVE HEATER

Features

 

When should I use an Objective Heater

You should use an objective heater, if you:

  • Perform Oil Immersion Imaging with a Top Stage Incubator
  • Use dipping lens with an upright Top Stage Incubator

Why should I use an Objective Heater?

Because, if you don't, sample temperature will drop due to oil contact with the cold objective, as shown in the following graph.

Step 1: Objective not in contact with Petri 2; Objective Heater Off.

Step 2: Objective in oil contact with Petri 2; Objective Heater Off.

Step 3: Objective in oil contact with Petri 2; Objective Heater on, Offset value 5.3°C.

During the self calibration routine, the offset value of the objective heater is progressively increased, until T2 reaches the desired set point temperature of 37.0°C.

Figure 3 reports the profile of T1, T2, and of Objective Heater Temperature during the experimental campaign described above.

 

 

 

 

Customer care is our top priority

CONTACT US

Okolab S.r.l.

FACTORY NAPOLI: Via A. Olivetti, 1 - 80078 Pozzuoli, NA, Italy

FACTORY ROVERETO: Via delle Zigherane - Edificio B08.01, 38068 Rovereto, TN, Italy

 

  • Luca Lanzaro Ph.D
  •   Phone: +39 081 8062624
  •   E-mail: info[at]oko-lab[dot]com

Okolab USA Inc.

417 Thorn St., Suite 309, Sewickley - PA 15143, USA

 

  • Lara Petrak
  • Mobile: +1 (650) 410 0756
  •   E-mail: usa[at]oko-lab[dot]com

Okolab Shanghai Co., Ltd.

Room 2604, Jiande Apartment, No. 9 Jiande Road, Huangpu District, Shanghai 200025, China

 

  • Pony Tong
  • Mobile: +86 18500508458
  •   E-mail: tong[at]oko-lab[dot]cn
 

COPYRIGHT © OKOLAB S.R.L. ALL RIGHT RESERVED. CERTIFIED ISO 9001.